The Kanomax tracer gas diffuser kit is used in performing tracer gas tests in accordance with ANSI / ASHRAE Standard 110. The diffuser is designed to the specifications of standard drawing #110-83M.
The diffuser is placed in the fume hood and sulfur hexafluoride gas is injected at at supply pressure of 30 psig. The internal critical orifice ensures a flow rate of 4 liters per minute. Other orifice sizes can be inserted.
Features & Benefits
Diffuser is made to the design and specifications of ASHRAE Standard Drawing #110-83M
The internal critical orifice ensures a flow rate of 4 liters per minute (lpm)
Thermal anemometer for face velocity testing
Commercial fog machine for flow visualization
Applications
Fume hood performance test according to ASHRAE Standard 110-2016
Main Unit Specifications
| Critical Orifice | 4 Liters Per Minute |
| Regulator | Dual Stage, Specialty Gas |
| Flow Meter | Calibrated 150mm/200psi, Glass Tube |
| Pressure Gauge | Large Dial, 0 to 60psi |
| Warranty | 1 year |
| Standard Accessories | Diffuser, Pressure Gauge, Shut-off Valve, In-line Flow Meter, 20 ft of Tubing, Tank Regulator for Gas Cylinder |
Hotline:08620-36453501